• Memmert ICH260C Climate Chamber [Resale]

    *Additional charge for installation
    *One Science does not handle the warranty for resale items


    Setting temperature range: -10 to +60 °C
    Working temperature range: from -10°C up to +60°C without humidity (not suitable for long-term storing at sub-zero temperatures. During permanent operation, the glass door may ice over)
    Working temperature range: from +10°C up to +60°C with humidity and/or light
    Setting accuracy temperature: 0.1 °C
    Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
    Works calibration certificate: standard values +10 °C and +37 °C, 60 % rh at +30 °C
    Door: inner glass doors
    Door: fully insulated stainless steel doors with 2-point locking (compression door lock)
    Entry port: Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed by a silicone stopper, standard position at the back
    Humidity: humidity supply with distilled water from external tank by self-priming pump
    Humidification: humidification by hot steam generator
    Humidity: active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity – resolution of display 0.1 %, setting accuracy 0.5 %
    Fan: forced air circulation
    External Dimensions: w(D) x h(E) x d(F): 824 x 1552 x 685 mm (d +56mm door handle)
    Internal Dimensions: w(A) x h(B) x d(C): 640 x 800 x 500 mm (d less 33 mm for fan)
    Volume: 256 l
    Max. number of internals: 9
    Max. loading of chamber: 200 kg
    Max. loading per internal: 20 kg

  • PB-30 pH Meter

    Measuring range of pH: -1.99 to 19.99

    pH readability: 0.01

    pH precision: ±0.005

    Measuring range of mV (mV): -1999.9 to 1999.9

    mV readability (mV): 0.1

    mV precision (mV): ±0.3 mV (0.05%, if +400 mV)

    Temperature range (°C): -5.0 to 105.0

    Readability (°C): 0.1

    Precision (°C): ±0.2

    Calibration point: Max. 3 buffers

    Automatic identification of buffers:

    16 buffers

    2, 4, 7, 10, 12

    1, 3, 6, 8, 10, 13

    1.68, 4.01, 6.86, 9.18, 12.46

    Electrode calibration slope range: 90% – 105%

    Temperature compensation: Automatic (ATC), manual (MTC)

    Optional measurement units: pH, mV and RmV (relative mV value)

     

    OPTIONAL ACCESSORIES:

  • PB-10 Standard pH Meter

    Measuring range of pH: 0 to 14

    pH readability: 0.01

    pH precision: ±0.01

    Measuring range of mV (mV): -1500.0~ to 1500.0

    mV readability (mV): 0.1

    mV precision (mV): ±0.4

    Temperature range (°C): -5.0 to 105.0

    Readability (°C): 0.1

    Precision (°C): ±0.2

    Calibration point: Max. 3 buffers

    Automatic identification of buffers:

    16 buffers

    2, 4, 7, 10, 12

    1, 3, 6,  8,  10, 13

    1.68, 4.01, 6.86, 9.18, 12.46

    Automatic temperature compensation (ATC): Yes

    Display of electrode slope: Yes

    Interface: BNC

    Display screen: LCD


    OPTIONAL ACCESSORIES:

Need Help?